Conductivity & PH Meter PDF Print E-mail
CPHT 003 - Device for measuring conductivity and pH
CPHT 003

The device is designed to measure conductivity and pH of solutions and solid substances;

  • super pure, drinking, river and sea water;
  • aqueous solutions of salts;
  • alcoholic solutions of salts;
  • acids, bases and other chemically active liquids by means of a special conductivity electrode;
  • milk, wine, soft carbonated and non-carbonated drinks;
  • pH of solid substances (e.g. cheese, yellow cheese, yoghurt, etc.) by means of a special pH electrode.
CPHT 003

Technical specifications

1. Conductivity measurement

1.1.Range: from 4,000 µs/cm to 4,000 s/cm in seven sub ranges.
1.2.Accuracy and resolution

Range 1S/cm* 100ms/cm 10mS/cm 1mS/cm 100µS/cm 10µS/cm 1µS/cm**
Accuracy ± 1% ± 0,5% ± 0,13% ± 0,13% ± 0,13% ± 0,5% ± 1%
Resolution 0,1 mS/cm 10 µS/cm 1 µS/cm 10 nS/cm 1 nS/cm 0,1 nS/cm 0,01 nS/cm

1.3. Option for automatic and manual change of the sub ranges
1.4. Thermal compensation - automatic, through in-built thermal sensor in the following modes:

- standard according to EN 27888
- linear with TKR = from 1 to 3,55% / °C
- switched off

1.5. Reference temperature - adjustable within the range from 0 to 99,9°C
1.6. Option for choice of one of 4 electrodes with constant values from 0,01 to 10,00 cm-1 for individual preservation of their constant values in calibration.
1.7. Calibration of the selected electrode - with any calibration solution of known conductivity within the range from 0,4 to 40 mS/cm at 25°C

2. Measuring pH

2.1. Range form 0,00 to 14,00 pH
2.2. Accuracy: ± 0,02 pH
2.3. Temperature compensation - automatic through separate thermal sensor
2.4. Calibration - with standard solutions of pH = 7,00; 4,00; 10,00
2.5. Measuring pH of solid substances - by means of a special electrode

3. Overall specifications:

3.1. Display - 4 x 16 sign/digital symbols with backlight
3.2. Power supply 4x1,5 V / 1,9 Ah alkaline rechargeable batteries and AC adapter
3.3. Option for simultaneous measurement of pH and conductivity
3.4. RS 232 interface for connection with PC and printer
3.5. Option for switching on and off the indication backlight
3.6. Adjustable automatic switch-off of the device
3.7. Charge and discharge control of the rechargeable battery
3.8. Compact design

User Guide Industrial pH Meter - 233 KB